Routine individualised patient dosimetry using electronic portal imaging devices.

Nijsten SM, Mijnheer BJ, Dekker AL, Lambin P, Minken AW

Department of Radiation Oncology (MAASTRO), GROW, University Hospital of Maastricht, Maastricht, The Netherlands. bas.nijsten@maastro.nl

BACKGROUND AND PURPOSE: To analyse the results of routine EPID measurements for individualised patient dosimetry. MATERIALS AND METHODS: Calibrated camera-based EPIDs were used to measure the central field dose, which was compared with a dose prediction at the EPID level. For transit dosimetry, dose data were calculated using patient transmission and scatter, and compared with measured values. Furthermore, measured transit dose data were back-projected to an in vivo dose value at 5 cm depth in water (D(5)) and directly compared with D(5) from the treatment planning system. Dose differences per treatment session were calculated by weighting dose values with the number of monitor units per beam. Reported errors were categorised and analysed for approximately 37,500 images from 2511 patients during a period of 24 months. RESULTS: Pre-treatment measurements showed a mean dose difference per treatment session of 0.0+/-1.7% (1 SD). Transfer errors were detected and corrected prior to the first treatment session. An accelerator output variation of about 4% was found between two weekly QC measurements. Patient dosimetry showed mean transit and D(5) dose differences of -0.7+/-5.2% (1 SD) and -0.3+/-5.6% (1 SD) per treatment session, respectively. Dose differences could be related to set-up errors, organ motion, erroneous density corrections and changes in patient anatomy. CONCLUSIONS: EPIDs can be used routinely to accurately verify treatment parameter transfer and machine output. By applying transit and in vivo dosimetry, more insight can be obtained with respect to the different error sources influencing dose delivery to a patient.

Published 9 April 2007 in Radiother Oncol, 83(1): 65-75.
Full-text of this article is available online (may require subscription).

Place a permanent text-link or advertisement here for just US$15.

© 2004-2008 Radiotherapy Research Today. All Rights Reserved.



Radiotherapy Research Today Archive:

Volume 1 (2004)
  Issue 1 (August)
  Issue 2 (September)
  Issue 3 (October)
  Issue 4 (November)
  Issue 5 (December)

Volume 2 (2005)
  Issue 1 (January)
  Issue 2 (February)
  Issue 3 (March)
  Issue 4 (April)
  Issue 5 (May)
  Issue 6 (June)
  Issue 7 (July)
  Issue 8 (August)
  Issue 9 (September)
  Issue 10 (October)
  Issue 11 (November)
  Issue 12 (December)

Volume 3 (2006)
  Issue 1 (January)
  Issue 2 (February)
  Issue 3 (March)
  Issue 4 (April)
  Issue 5 (May)
  Issue 6 (June)
  Issue 7 (July)
  Issue 8 (August)
  Issue 9 (September)
  Issue 10 (October)
  Issue 11 (November)
  Issue 12 (December)

Volume 4 (2007)
  Issue 1 (January)
  Issue 2 (February)
  Issue 3 (March)
  Issue 4 (April)
  Issue 5 (May)
  Issue 6 (June)
  Issue 7 (July)
  Issue 8 (August)
  Issue 9 (September)
  Issue 10 (October)
  Issue 11 (November)
  Issue 12 (December)

Volume 5 (2008)
  Issue 1 (January)
  Issue 2 (February)
  Issue 3 (March)
  Issue 4 (April)
  Issue 5 (May)
  Issue 6 (June)
  Issue 7 (July)
  Issue 8 (August)
  Issue 9 (September)



Radiotherapy Books

Radiation Oncology for Cure and Palliation(Medical Radiology / Radiation Oncology)

Radiation Oncology for Cure and Palliation(Medical Radiology / Radiation Oncology)