Positioning accuracy of cone-beam computed tomography in combination with a HexaPOD robot treatment table.

Meyer J, Wilbert J, Baier K, Guckenberger M, Richter A, Sauer O, Flentje M

Department of Physics and Astronomy, University of Canterbury, Christchurch, New Zealand. juergen.meyer@canterbury.ac.nz

PURPOSE: To scrutinize the positioning accuracy and reproducibility of a commercial hexapod robot treatment table (HRTT) in combination with a commercial cone-beam computed tomography system for image-guided radiotherapy (IGRT). METHODS AND MATERIALS: The mechanical stability of the X-ray volume imaging (XVI) system was tested in terms of reproducibility and with a focus on the moveable parts, i.e., the influence of kV panel and the source arm on the reproducibility and accuracy of both bone and gray value registration using a head-and-neck phantom. In consecutive measurements the accuracy of the HRTT for translational, rotational, and a combination of translational and rotational corrections was investigated. The operational range of the HRTT was also determined and analyzed. RESULTS: The system performance of the XVI system alone was very stable with mean translational and rotational errors of below 0.2 mm and below 0.2 degrees , respectively. The mean positioning accuracy of the HRTT in combination with the XVI system summarized over all measurements was below 0.3 mm and below 0.3 degrees for translational and rotational corrections, respectively. The gray value match was more accurate than the bone match. CONCLUSION: The XVI image acquisition and registration procedure were highly reproducible. Both translational and rotational positioning errors can be corrected very precisely with the HRTT. The HRTT is therefore well suited to complement cone-beam computed tomography to take full advantage of position correction in six degrees of freedom for IGRT. The combination of XVI and the HRTT has the potential to improve the accuracy of high-precision treatments.

Published 5 March 2007 in Int J Radiat Oncol Biol Phys, 67(4): 1220-8.
Full-text of this article is available online (may require subscription).

Place a permanent text-link or advertisement here for just US$15.

© 2004-2008 Radiotherapy Research Today. All Rights Reserved.



Radiotherapy Research Today Archive:

Volume 1 (2004)
  Issue 1 (August)
  Issue 2 (September)
  Issue 3 (October)
  Issue 4 (November)
  Issue 5 (December)

Volume 2 (2005)
  Issue 1 (January)
  Issue 2 (February)
  Issue 3 (March)
  Issue 4 (April)
  Issue 5 (May)
  Issue 6 (June)
  Issue 7 (July)
  Issue 8 (August)
  Issue 9 (September)
  Issue 10 (October)
  Issue 11 (November)
  Issue 12 (December)

Volume 3 (2006)
  Issue 1 (January)
  Issue 2 (February)
  Issue 3 (March)
  Issue 4 (April)
  Issue 5 (May)
  Issue 6 (June)
  Issue 7 (July)
  Issue 8 (August)
  Issue 9 (September)
  Issue 10 (October)
  Issue 11 (November)
  Issue 12 (December)

Volume 4 (2007)
  Issue 1 (January)
  Issue 2 (February)
  Issue 3 (March)
  Issue 4 (April)
  Issue 5 (May)
  Issue 6 (June)
  Issue 7 (July)
  Issue 8 (August)
  Issue 9 (September)
  Issue 10 (October)
  Issue 11 (November)
  Issue 12 (December)

Volume 5 (2008)
  Issue 1 (January)
  Issue 2 (February)
  Issue 3 (March)
  Issue 4 (April)
  Issue 5 (May)
  Issue 6 (June)
  Issue 7 (July)
  Issue 8 (August)



Radiotherapy Books

A Guide for Delineation of Lymph Nodal Clinical Target Volume in Radiation Therapy

A Guide for Delineation of Lymph Nodal Clinical Target Volume in Radiation Therapy