Radiotherapy Research Today is a free monthly online journal that collates and summarizes the latest research about Radiotherapy, including details on cancer treatment, side effects. | ||||||||
|
A 3D photon superposition/convolution algorithm and its foundation on results of Monte Carlo calculations.Ulmer W, Pyyry J, Kaissl W VARIAN Medical Systems International AG, Baden-Daettwil, Switzerland. waldemar.ulmer@varian.com Based on previous publications on a triple Gaussian analytical pencil beam model and on Monte Carlo calculations using Monte Carlo codes GEANT-Fluka, versions 95, 98, 2002, and BEAMnrc/EGSnrc, a three-dimensional (3D) superposition/convolution algorithm for photon beams (6 MV, 18 MV) is presented. Tissue heterogeneity is taken into account by electron density information of CT images. A clinical beam consists of a superposition of divergent pencil beams. A slab-geometry was used as a phantom model to test computed results by measurements. An essential result is the existence of further dose build-up and build-down effects in the domain of density discontinuities. These effects have increasing magnitude for field sizes < or =5.5 cm(2) and densities < or = 0.25 g cm(-3), in particular with regard to field sizes considered in stereotaxy. They could be confirmed by measurements (mean standard deviation 2%). A practical impact is the dose distribution at transitions from bone to soft tissue, lung or cavities. Published 7 April 2005 in Phys Med Biol, 50(8): 1767-90.